The '''quadrupole mass analyzer''' ('''QMS'''), also known as a '''transmission quadrupole mass spectrometer''', '''quadrupole mass filter''', or '''quadrupole mass spectrometer''', is one type of mass analyzer used in [[mass spectrometry]]. As the name implies, it consists of four cylindrical rods, set parallel to each other.<ref name="MSPA">{{cite book|last=de Hoffmann|first=Edmond|author2=Vincent Stroobant |title=Mass Spectrometry: Principles and Applications|publisher=John Wiley & Sons, Ltd.|location=Toronto|date=2003|edition=Second| isbn = 978-0-471-48566-7|page=65}}</ref> In a quadrupole [[mass spectrometer]] the [[quadrupole]] is the ''mass analyzer'' - the component of the instrument responsible for selecting sample ions based on their [[mass-to-charge ratio]] (''m/z''). Ions are separated in a quadrupole based on the stability of their trajectories in the oscillating [[electric field]]s that are applied to the rods.<ref name="MSPA" />
In [[mass spectrometry]], the '''quadrupole mass analyzer''' (or '''quadrupole mass filter''') is a type of [[mass]] analyzer originally conceived<ref name=":0" /> by Nobel laureate [[Wolfgang Paul]] and his student Helmut Steinwedel. As the name implies, it consists of four cylindrical rods, set parallel to each other.<ref name="MSPA">{{cite book|last=de Hoffmann|first=Edmond|author2=Vincent Stroobant |title=Mass Spectrometry: Principles and Applications|publisher=John Wiley & Sons, Ltd.|location=Toronto|date=2003|edition=Second| isbn = 978-0-471-48566-7|page=65}}</ref> In a quadrupole mass spectrometer (QMS) the [[quadrupole]] is the ''mass analyzer'' the component of the instrument responsible for selecting sample [[ion]]s based on their [[mass-to-charge ratio]] (''m/z''). Ions are separated in a quadrupole based on the stability of their trajectories in the oscillating [[electric field]]s that are applied to the rods.<ref name="MSPA" />


==Principle of operation==
==Principle of operation==
[[File:Paul Patent 2939952 Fig5.gif|thumb|right|400px|Image from "Apparatus For Separating Charged Particles Of Different Specific Charges" Patent number: 2939952<ref>{{Cite patent|US|2939952}} ''Apparatus For Separating Charged Particles Of Different Specific Charges'' Jun 1960; Paul et al.</ref>]] The quadrupole consists of four parallel metal rods. Each opposing rod pair is connected together electrically, and a [[radio frequency]] (RF) voltage with a DC offset voltage is applied between one pair of rods and the other. Ions travel down the quadrupole between the rods. Only ions of a certain mass-to-charge ratio will reach the detector for a given ratio of voltages: other ions have unstable trajectories and will collide with the rods. This permits selection of an ion with a particular ''m/z'' or allows the operator to scan for a range of ''m/z''-values by continuously varying the applied voltage.<ref name="MSPA" /> Mathematically this can be modeled with the help of the [[Mathieu differential equation]].<ref>{{cite book |author=Gerald Teschl |author-link=Gerald Teschl | title = Ordinary Differential Equations and Dynamical Systems | publisher=[[American Mathematical Society]] | place = [[Providence, Rhode Island|Providence]] | date = 2012 | isbn = 978-0-8218-8328-0 | url = https://www.mat.univie.ac.at/~gerald/ftp/book-ode/}}</ref>
[[File:Paul Patent 2939952 Fig5.gif|thumb|right|400px|Image from US Patent "Apparatus for separating charged particles of different specific charges"<ref name=":0">{{Cite patent|country=US|number=2939952|pubdate=1960-06-07|title=Apparatus for separating charged particles of different specific charges|assign1=Wolfgang Paul|inventor1-last=Paul|inventor1-first=Wolfgang|inventor2-last=Steinwedel|inventor2-first=Helmut|inventorlink1=Wolfgang_Paul}}</ref>]] The quadrupole consists of four parallel metal rods. Each opposing rod pair is connected together electrically, and a [[radio frequency]] (RF) voltage with a DC offset voltage is applied between one pair of rods and the other. Ions travel down the quadrupole between the rods. Only ions of a certain mass-to-charge ratio will reach the detector for a given ratio of voltages: other ions have unstable trajectories and will collide with the rods. This permits selection of an ion with a particular ''m/z'' or allows the operator to scan for a range of ''m/z''-values by continuously varying the applied voltage.<ref name="MSPA" /> Mathematically this can be modeled with the help of the [[Mathieu differential equation]].<ref>{{cite book |author=Gerald Teschl |author-link=Gerald Teschl | title = Ordinary Differential Equations and Dynamical Systems | publisher=[[American Mathematical Society]] | place = [[Providence, Rhode Island|Providence]] | date = 2012 | isbn = 978-0-8218-8328-0 | url = https://www.mat.univie.ac.at/~gerald/ftp/book-ode/}}</ref>


[[File:Quadrupole mass analyzer.svg|thumb|right|400px|Ion path through a quadrupole]]
[[File:Quadrupole mass analyzer.svg|thumb|right|400px|Ion path through a quadrupole]]


Ideally, the rods are [[hyperbola|hyperbolic]]. Cylindrical rods with a specific ratio of rod diameter-to-spacing provide an easier-to-manufacture adequate approximation to hyperbolas. Small variations in the ratio have large effects on resolution and peak shape. Different manufacturers choose slightly different ratios to fine-tune operating characteristics in context of anticipated application requirements. In recent decades some manufacturers have produced quadrupole mass spectrometers with true hyperbolic rods.
Ideally, the rods are [[hyperbola|hyperbolic]], however cylindrical rods with a specific ratio of rod diameter-to-spacing provide an easier-to-manufacture adequate approximation to hyperbolas. Small variations in the ratio have large effects on resolution and peak shape. Different manufacturers choose slightly different ratios to fine-tune operating characteristics in context of anticipated application requirements. Since the 1980s, the [[Atlas Werke|MAT]] company and subsequently [[Finnigan Instrument Corporation]] used hyperbolic rods produced with a mechanical tolerance of 0.001 mm, whose exact production process was a well-kept secret within the company.<ref name="50years">{{Cite journal|url=https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/abs/10.1002/%28SICI%291097-0231%28199704%2911%3A6%3C694%3A%3AAID-RCM888%3E3.0.CO%3B2-K|title=50 Years of MAT in Bremen|first=Curt|last=Brunnée|date=May 27, 1997|journal=Rapid Communications in Mass Spectrometry|volume=11|issue=6|pages=694–707|via=Wiley Online Library|doi=10.1002/(SICI)1097-0231(199704)11:6<694::AID-RCM888>3.0.CO;2-K}}</ref>


==Multiple quadrupoles, hybrids and variations==
==Multiple quadrupoles, hybrids and variations==
Line 14: Line 16:
A linear series of three quadrupoles is known as a [[triple quadrupole mass spectrometer]]. The first (Q<sub>1</sub>) and third (Q<sub>3</sub>) quadrupoles act as mass filters, and the middle (q<sub>2</sub>) quadrupole is employed as a collision cell. This collision cell is an RF-only quadrupole (non-mass filtering) using Ar, He, or N<sub>2</sub> gas (~10<sup>−3</sup> Torr, ~30 eV) for collision induced dissociation of selected parent ion(s) from Q<sub>1</sub>. Subsequent fragments are passed through to Q<sub>3</sub> where they may be filtered or fully scanned.
A linear series of three quadrupoles is known as a [[triple quadrupole mass spectrometer]]. The first (Q<sub>1</sub>) and third (Q<sub>3</sub>) quadrupoles act as mass filters, and the middle (q<sub>2</sub>) quadrupole is employed as a collision cell. This collision cell is an RF-only quadrupole (non-mass filtering) using Ar, He, or N<sub>2</sub> gas (~10<sup>−3</sup> Torr, ~30 eV) for collision induced dissociation of selected parent ion(s) from Q<sub>1</sub>. Subsequent fragments are passed through to Q<sub>3</sub> where they may be filtered or fully scanned.


This process allows for the study of fragments that are useful in structural elucidation by [[tandem mass spectrometry]]. For example, the Q<sub>1</sub> may be set to 'filter' for a drug ion of known mass, which is fragmented in q<sub>2</sub>. The third quadrupole (Q<sub>3</sub>) can then be set to scan the entire ''m/z'' range, giving information on the intensities of the fragments. Thus, the structure of the original ion can be deduced.
This process allows for the study of fragments that are useful in [[structural elucidation]] by [[tandem mass spectrometry]]. For example, the Q<sub>1</sub> may be set to 'filter' for a drug ion of known mass, which is fragmented in q<sub>2</sub>. The third quadrupole (Q<sub>3</sub>) can then be set to scan the entire ''m/z'' range, giving information on the intensities of the fragments. Thus, the structure of the original ion can be deduced.


The arrangement of three quadrupoles was first developed by Jim Morrison of LaTrobe University, Australia for the purpose of studying the [[photodissociation]] of gas-phase ions.<ref>{{Cite journal | last = Morrison | first = J. D. | date = 1991 | title = Personal reminiscences of forty years of mass spectrometry in Australia | journal = Organic Mass Spectrometry | volume = 26 | pages = 183–194 | doi = 10.1002/oms.1210260404 | issue = 4}}</ref> The first triple-quadrupole mass spectrometer was developed at Michigan State University by [[Christie G. Enke|Christie Enke]] and graduate student [[Richard Yost]] in the late 1970s.<ref>{{Cite journal | last = Yost | first = R. A. | date = 1978 | title = Selected ion fragmentation with a tandem quadrupole mass spectrometer | journal = Journal of the American Chemical Society | volume = 100 | pages = 2274 | doi = 10.1021/ja00475a072 | url = http://masspec.scripps.edu/mshistory/timeline/time_pdf/1978_YostRA.pdf | last2 = Enke | first2 = C. G. | issue = 7 | access-date = 2008-12-06 | archive-url=https://web.archive.org/web/20120219050825/http://masspec.scripps.edu/mshistory/timeline/time_pdf/1978_YostRA.pdf | archive-date = 2012-02-19 | url-status = dead}}</ref>
The arrangement of three quadrupoles was first developed by [[Jim Morrison (chemist)|Jim Morrison]] of [[La Trobe University]] in Australia for the purpose of studying the [[photodissociation]] of gas-phase ions.<ref>{{Cite journal | last = Morrison | first = J. D. | date = 1991 | title = Personal reminiscences of forty years of mass spectrometry in Australia | journal = Organic Mass Spectrometry | volume = 26 | pages = 183–194 | doi = 10.1002/oms.1210260404 | issue = 4}}</ref> The first triple-quadrupole mass spectrometer was developed at Michigan State University by [[Christie G. Enke|Christie Enke]] and graduate student [[Richard Yost]] in the late 1970s.<ref>{{Cite journal | last1 = Yost | first1 = R. A. | date = 1978 | title = Selected ion fragmentation with a tandem quadrupole mass spectrometer | journal = Journal of the American Chemical Society | volume = 100 | pages = 2274 | doi = 10.1021/ja00475a072 | url = http://masspec.scripps.edu/mshistory/timeline/time_pdf/1978_YostRA.pdf | last2 = Enke | first2 = C. G. | issue = 7 | access-date = 2008-12-06 | archive-url=https://web.archive.org/web/20120219050825/http://masspec.scripps.edu/mshistory/timeline/time_pdf/1978_YostRA.pdf | archive-date = 2012-02-19 | url-status = dead}}</ref>


Quadrupoles can be used in [[hybrid mass spectrometer]]s. For example, a [[sector instrument]] can be combined with a collision quadrupole and quadrupole mass analyzer to form a hybrid instrument.<ref>{{Cite journal | last = Glish | first = G. | date = 1982 | title = A new "hybrid" sector/quadrupole mass spectrometer for mass spectrometry/mass spectrometry | journal = International Journal of Mass Spectrometry and Ion Physics | volume = 41 | pages = 157 | doi = 10.1016/0020-7381(82)85032-8 |author2=Scott A. McLuckey |author2-link=Scott A. McLuckey | last3 = Ridley | first3 = T | last4 = Cooks | first4 = R | postscript = <!--None--> | issue = 3|bibcode = 1982IJMSI..41..157G}}</ref>
Quadrupoles can be used in [[hybrid mass spectrometer]]s. For example, a [[sector instrument]] can be combined with a collision quadrupole and quadrupole mass analyzer to form a hybrid instrument.<ref>{{Cite journal | last = Glish | first = G. | date = 1982 | title = A new "hybrid" sector/quadrupole mass spectrometer for mass spectrometry/mass spectrometry | journal = International Journal of Mass Spectrometry and Ion Physics | volume = 41 | pages = 157 | doi = 10.1016/0020-7381(82)85032-8 |author2=Scott A. McLuckey |author2-link=Scott A. McLuckey | last3 = Ridley | first3 = T | last4 = Cooks | first4 = R | issue = 3|bibcode = 1982IJMSI..41..157G}}</ref>


A mass selecting quadrupole and collision quadrupole with [[time-of-flight]] device as the second mass selection stage is a hybrid known as a [[quadrupole time-of-flight mass spectrometer]] (QTOF MS).<ref>{{cite journal |author=Shevchenko A |author2=Loboda A |author3=Shevchenko A|author4=Ens W |author5=Standing KG |title=MALDI quadrupole time-of-flight mass spectrometry: a powerful tool for proteomic research |journal=Anal. Chem. |volume=72 |issue=9 |pages=2132–41 |date=May 2000 |pmid=10815976 |doi= 10.1021/ac9913659}}</ref><ref>{{cite journal |author=Steen H |author2=Küster B |author3= Mann M |title=Quadrupole time-of-flight versus triple-quadrupole mass spectrometry for the determination of phosphopeptides by precursor ion scanning |journal=J Mass Spectrom |volume=36 |issue=7 |pages=782–90 |date=July 2001 |pmid=11473401 |doi=10.1002/jms.174|bibcode=2001JMSp...36..782S}}</ref> QTOFs are used for the mass spectrometry of peptides and other large biological polymers.<ref>{{Cite journal | last = Chernushevich | first = Igor V. | date = 2001 | title = An introduction to quadrupole–time-of-flight mass spectrometry | journal = Journal of Mass Spectrometry | volume = 36 | pages = 849–865 | issue = 8 | doi=10.1002/jms.207| pmid = 11523084 | bibcode = 2001JMSp...36..849C}}</ref>
A mass-selecting quadrupole and collision quadrupole with [[time-of-flight]] device as the second mass selection stage is a hybrid known as a [[quadrupole time-of-flight mass spectrometer]] (QTOF MS).<ref>{{cite journal |author=Shevchenko A |author2=Loboda A |author3=Shevchenko A|author4=Ens W |author5=Standing KG |title=MALDI quadrupole time-of-flight mass spectrometry: a powerful tool for proteomic research |journal=Anal. Chem. |volume=72 |issue=9 |pages=2132–41 |date=May 2000 |pmid=10815976 |doi= 10.1021/ac9913659}}</ref><ref>{{cite journal |author=Steen H |author2=Küster B |author3= Mann M |title=Quadrupole time-of-flight versus triple-quadrupole mass spectrometry for the determination of phosphopeptides by precursor ion scanning |journal=J Mass Spectrom |volume=36 |issue=7 |pages=782–90 |date=July 2001 |pmid=11473401 |doi=10.1002/jms.174|bibcode=2001JMSp...36..782S}}</ref> Quadrupole-quadrupole-time-of-flight (QqTOF) configurations are also possible and used especially the mass spectrometry of peptides and other large biological polymers.<ref>{{Cite journal | last = Chernushevich | first = Igor V. | date = 2001 | title = An introduction to quadrupole–time-of-flight mass spectrometry | journal = Journal of Mass Spectrometry | volume = 36 | pages = 849–865 | issue = 8 | doi=10.1002/jms.207| pmid = 11523084 | bibcode = 2001JMSp...36..849C}}</ref><ref>{{Cite journal|last1=Oberacher|first1=Herbert|last2=Pitterl|first2=Florian|date=June 2009|editor-last=Fabris|editor-first=Dan|title=On the use of ESI-QqTOF-MS/MS for the comparative sequencing of nucleic acids|url=http://doi.wiley.com/10.1002/bip.21156|journal=Biopolymers|language=en|volume=91|issue=6|pages=401–409|doi=10.1002/bip.21156|pmid=19189378 }}</ref>


A variant of the quadrupole mass analyzer called the monopole was invented by von Zahn which operates with two electrodes and generates one quarter of the quadrupole field.<ref>{{cite journal |author=U. von Zahn|title=Monopole Spectrometer, a New Electric Field Mass Spectrometer |journal=Rev. Sci. Instrum. |volume=34 |issue=1 |pages=1–4 |date=1963 |doi=10.1063/1.1718110|bibcode = 1963RScI...34....1V}}</ref> It has one circular electrode and one ''v'' shaped electrode. The performance is however lower than of the quadrupole mass analyzer.
A variant of the quadrupole mass analyzer called the monopole was invented by von Zahn which operates with two electrodes and generates one quarter of the quadrupole field.<ref>{{cite journal |author=U. von Zahn|title=Monopole Spectrometer, a New Electric Field Mass Spectrometer |journal=Rev. Sci. Instrum. |volume=34 |issue=1 |pages=1–4 |date=1963 |doi=10.1063/1.1718110|bibcode = 1963RScI...34....1V}}</ref> It has one circular electrode and one V-shaped electrode. The performance is, however, lower than that of a quadrupole mass analyzer.


An enhancement to the performance of the quadrupole mass analyzer has been demonstrated to occur when a magnetic field is applied to the instrument. Manyfold improvements in resolution and sensitivity have been reported for a magnetic field applied in various orientations to a QMS.<ref>{{cite journal |author=Syed S. |author2=Maher S. |author3=Taylor S. |title=Quadrupole mass filter operation under the influence of magnetic field |journal=Journal of Mass Spectrometry |volume=48 |issue=12 |pages=1325–1339 |date=2013 |doi= 10.1002/jms.3293|pmid=24338888|bibcode=2013JMSp...48.1325S}}</ref><ref>{{cite journal |author=Maher S |author2=Syed S |author3=Hughes D|author4=Gibson J |author5=Taylor S |title=Mapping the stability diagram of a quadrupole mass spectrometer with a static transverse magnetic field applied |journal=Journal of the American Society for Mass Spectrometry |volume=24 |issue=8 |pages=1307–1314 |date=2013 |doi=10.1007/s13361-013-0654-5|pmid=23720050 |bibcode = 2013JASMS..24.1307M}}</ref>
An enhancement to the performance of the quadrupole mass analyzer has been demonstrated to occur when a magnetic field is applied to the instrument. Manifold improvements in resolution and sensitivity have been reported for a magnetic field applied in various orientations to a QMS.<ref>{{cite journal |author=Syed S. |author2=Maher S. |author3=Taylor S. |title=Quadrupole mass filter operation under the influence of magnetic field |journal=Journal of Mass Spectrometry |volume=48 |issue=12 |pages=1325–1339 |date=2013 |doi= 10.1002/jms.3293|pmid=24338888|bibcode=2013JMSp...48.1325S}}</ref><ref>{{cite journal |author=Maher S |author2=Syed S |author3=Hughes D|author4=Gibson J |author5=Taylor S |title=Mapping the stability diagram of a quadrupole mass spectrometer with a static transverse magnetic field applied |journal=Journal of the American Society for Mass Spectrometry |volume=24 |issue=8 |pages=1307–1314 |date=2013 |doi=10.1007/s13361-013-0654-5|pmid=23720050 |bibcode = 2013JASMS..24.1307M|s2cid=45734248 }}</ref>


==Applications==
==Applications==
These mass spectrometers excel at applications where particular ions of interest are being studied because they can stay tuned on a single ion for extended periods of time. One place where this is useful is in [[liquid chromatography-mass spectrometry]] or [[gas chromatography-mass spectrometry]] where they serve as exceptionally high specificity detectors. Quadrupole instruments are often reasonably priced and make good multi-purpose instruments. The single quadrupole mass spectrometer with electron impact ionizer is used as a standalone analyser in [[residual gas analyzer]]s, real time gas analyzers, [[plasma diagnostics]] and SIMS surface analysis systems.
These mass spectrometers excel at applications where particular ions of interest are being studied because they can stay tuned on a single ion for extended periods of time. One place where this is useful is in [[liquid chromatography-mass spectrometry]] or [[gas chromatography-mass spectrometry]] where they serve as exceptionally high specificity detectors. Quadrupole instruments are often reasonably priced and make good multi-purpose instruments. A single quadrupole mass spectrometer with an electron impact ionizer is used as a standalone analyzer in [[residual gas analyzer]]s, real-time gas analyzers, [[plasma diagnostics]] and [[Secondary ion mass spectrometry|SIMS surface analysis]] systems.{{Citation needed|date=November 2021}}


==See also==
==See also==
* [[Fourier transform ion cyclotron resonance]]
* [[Quadrupole]]
* [[Quadrupole magnet]]
* [[Quadrupole magnet]]
* [[RFQ beam cooler|Radio-frequency quadrupole beam cooler]]
* [[RFQ beam cooler|Radio-frequency quadrupole beam cooler]]
* [[Fourier transform ion cyclotron resonance]]


==References==
==References==

Latest revision as of 17:26, 27 December 2023

Quadrupole elements

In mass spectrometry, the quadrupole mass analyzer (or quadrupole mass filter) is a type of mass analyzer originally conceived[1] by Nobel laureate Wolfgang Paul and his student Helmut Steinwedel. As the name implies, it consists of four cylindrical rods, set parallel to each other.[2] In a quadrupole mass spectrometer (QMS) the quadrupole is the mass analyzer – the component of the instrument responsible for selecting sample ions based on their mass-to-charge ratio (m/z). Ions are separated in a quadrupole based on the stability of their trajectories in the oscillating electric fields that are applied to the rods.[2]

Principle of operation

[edit]
Image from US Patent "Apparatus for separating charged particles of different specific charges"[1]

The quadrupole consists of four parallel metal rods. Each opposing rod pair is connected together electrically, and a radio frequency (RF) voltage with a DC offset voltage is applied between one pair of rods and the other. Ions travel down the quadrupole between the rods. Only ions of a certain mass-to-charge ratio will reach the detector for a given ratio of voltages: other ions have unstable trajectories and will collide with the rods. This permits selection of an ion with a particular m/z or allows the operator to scan for a range of m/z-values by continuously varying the applied voltage.[2] Mathematically this can be modeled with the help of the Mathieu differential equation.[3]

Ion path through a quadrupole

Ideally, the rods are hyperbolic, however cylindrical rods with a specific ratio of rod diameter-to-spacing provide an easier-to-manufacture adequate approximation to hyperbolas. Small variations in the ratio have large effects on resolution and peak shape. Different manufacturers choose slightly different ratios to fine-tune operating characteristics in context of anticipated application requirements. Since the 1980s, the MAT company and subsequently Finnigan Instrument Corporation used hyperbolic rods produced with a mechanical tolerance of 0.001 mm, whose exact production process was a well-kept secret within the company.[4]

Multiple quadrupoles, hybrids and variations

[edit]
Hybrid quadrupole time-of-flight mass spectrometer

A linear series of three quadrupoles is known as a triple quadrupole mass spectrometer. The first (Q1) and third (Q3) quadrupoles act as mass filters, and the middle (q2) quadrupole is employed as a collision cell. This collision cell is an RF-only quadrupole (non-mass filtering) using Ar, He, or N2 gas (~10−3 Torr, ~30 eV) for collision induced dissociation of selected parent ion(s) from Q1. Subsequent fragments are passed through to Q3 where they may be filtered or fully scanned.

This process allows for the study of fragments that are useful in structural elucidation by tandem mass spectrometry. For example, the Q1 may be set to 'filter' for a drug ion of known mass, which is fragmented in q2. The third quadrupole (Q3) can then be set to scan the entire m/z range, giving information on the intensities of the fragments. Thus, the structure of the original ion can be deduced.

The arrangement of three quadrupoles was first developed by Jim Morrison of La Trobe University in Australia for the purpose of studying the photodissociation of gas-phase ions.[5] The first triple-quadrupole mass spectrometer was developed at Michigan State University by Christie Enke and graduate student Richard Yost in the late 1970s.[6]

Quadrupoles can be used in hybrid mass spectrometers. For example, a sector instrument can be combined with a collision quadrupole and quadrupole mass analyzer to form a hybrid instrument.[7]

A mass-selecting quadrupole and collision quadrupole with time-of-flight device as the second mass selection stage is a hybrid known as a quadrupole time-of-flight mass spectrometer (QTOF MS).[8][9] Quadrupole-quadrupole-time-of-flight (QqTOF) configurations are also possible and used especially the mass spectrometry of peptides and other large biological polymers.[10][11]

A variant of the quadrupole mass analyzer called the monopole was invented by von Zahn which operates with two electrodes and generates one quarter of the quadrupole field.[12] It has one circular electrode and one V-shaped electrode. The performance is, however, lower than that of a quadrupole mass analyzer.

An enhancement to the performance of the quadrupole mass analyzer has been demonstrated to occur when a magnetic field is applied to the instrument. Manifold improvements in resolution and sensitivity have been reported for a magnetic field applied in various orientations to a QMS.[13][14]

Applications

[edit]

These mass spectrometers excel at applications where particular ions of interest are being studied because they can stay tuned on a single ion for extended periods of time. One place where this is useful is in liquid chromatography-mass spectrometry or gas chromatography-mass spectrometry where they serve as exceptionally high specificity detectors. Quadrupole instruments are often reasonably priced and make good multi-purpose instruments. A single quadrupole mass spectrometer with an electron impact ionizer is used as a standalone analyzer in residual gas analyzers, real-time gas analyzers, plasma diagnostics and SIMS surface analysis systems.[citation needed]

See also

[edit]

References

[edit]
  1. ^ a b US 2939952, Paul, Wolfgang & Steinwedel, Helmut, "Apparatus for separating charged particles of different specific charges", published 1960-06-07, assigned to Wolfgang Paul 
  2. ^ a b c de Hoffmann, Edmond; Vincent Stroobant (2003). Mass Spectrometry: Principles and Applications (Second ed.). Toronto: John Wiley & Sons, Ltd. p. 65. ISBN 978-0-471-48566-7.
  3. ^ Gerald Teschl (2012). Ordinary Differential Equations and Dynamical Systems. Providence: American Mathematical Society. ISBN 978-0-8218-8328-0.
  4. ^ Brunnée, Curt (May 27, 1997). "50 Years of MAT in Bremen". Rapid Communications in Mass Spectrometry. 11 (6): 694–707. doi:10.1002/(SICI)1097-0231(199704)11:6<694::AID-RCM888>3.0.CO;2-K – via Wiley Online Library.
  5. ^ Morrison, J. D. (1991). "Personal reminiscences of forty years of mass spectrometry in Australia". Organic Mass Spectrometry. 26 (4): 183–194. doi:10.1002/oms.1210260404.
  6. ^ Yost, R. A.; Enke, C. G. (1978). "Selected ion fragmentation with a tandem quadrupole mass spectrometer" (PDF). Journal of the American Chemical Society. 100 (7): 2274. doi:10.1021/ja00475a072. Archived from the original (PDF) on 2012-02-19. Retrieved 2008-12-06.
  7. ^ Glish, G.; Scott A. McLuckey; Ridley, T; Cooks, R (1982). "A new "hybrid" sector/quadrupole mass spectrometer for mass spectrometry/mass spectrometry". International Journal of Mass Spectrometry and Ion Physics. 41 (3): 157. Bibcode:1982IJMSI..41..157G. doi:10.1016/0020-7381(82)85032-8.
  8. ^ Shevchenko A; Loboda A; Shevchenko A; Ens W; Standing KG (May 2000). "MALDI quadrupole time-of-flight mass spectrometry: a powerful tool for proteomic research". Anal. Chem. 72 (9): 2132–41. doi:10.1021/ac9913659. PMID 10815976.
  9. ^ Steen H; Küster B; Mann M (July 2001). "Quadrupole time-of-flight versus triple-quadrupole mass spectrometry for the determination of phosphopeptides by precursor ion scanning". J Mass Spectrom. 36 (7): 782–90. Bibcode:2001JMSp...36..782S. doi:10.1002/jms.174. PMID 11473401.
  10. ^ Chernushevich, Igor V. (2001). "An introduction to quadrupole–time-of-flight mass spectrometry". Journal of Mass Spectrometry. 36 (8): 849–865. Bibcode:2001JMSp...36..849C. doi:10.1002/jms.207. PMID 11523084.
  11. ^ Oberacher, Herbert; Pitterl, Florian (June 2009). Fabris, Dan (ed.). "On the use of ESI-QqTOF-MS/MS for the comparative sequencing of nucleic acids". Biopolymers. 91 (6): 401–409. doi:10.1002/bip.21156. PMID 19189378.
  12. ^ U. von Zahn (1963). "Monopole Spectrometer, a New Electric Field Mass Spectrometer". Rev. Sci. Instrum. 34 (1): 1–4. Bibcode:1963RScI...34....1V. doi:10.1063/1.1718110.
  13. ^ Syed S.; Maher S.; Taylor S. (2013). "Quadrupole mass filter operation under the influence of magnetic field". Journal of Mass Spectrometry. 48 (12): 1325–1339. Bibcode:2013JMSp...48.1325S. doi:10.1002/jms.3293. PMID 24338888.
  14. ^ Maher S; Syed S; Hughes D; Gibson J; Taylor S (2013). "Mapping the stability diagram of a quadrupole mass spectrometer with a static transverse magnetic field applied". Journal of the American Society for Mass Spectrometry. 24 (8): 1307–1314. Bibcode:2013JASMS..24.1307M. doi:10.1007/s13361-013-0654-5. PMID 23720050. S2CID 45734248.